PAPP-A Test Kit – CLIA
PAPP-A Test Kit – ELISA

| Description: |
PAPP-A (Pregnancy Associated Plasma Protein A) |
|
| Catalog number: | AnshLite™ PAPP-A CLIA, AL-206 | |
| Kit / type: | 96-well (12×8 removable strips) microtiter plate CLIA | |
| Detection: | HRP-based enhanced chemiluminescence, utilizing our proprietary AnshLite™ luminogenic substrate | |
| Intended use: | Unless otherwise stated in our catalog or other company documentation accompanying Ansh Labs product(s), the products are intended for research use only, and not for in vitro diagnostic purposes, or therapeutic uses of any type involving consumption or application to humans or animals. | |
| Assay note(s): | Quantitative two-step sandwich type immunoassay. Apprpriate sample dilution should be performed same days as assay. Total 3 hour incubation time at room temperature. | |
| Sample type(s): | Serum, diluted 1:100 | |
| Sample size: | 50 μL of diluted specimens, 10 μL specimen into 990 μL assay buffer | |
| Analytical Specificity: | The antibody pair used in the PAPP-A kits measures dimeric PAPP-A and PAPP-A/proMBP complex approximately in equimolar concentration and does not cross-react with PAPP-A2 and MMP-9 when tested at 50 ng/mL. | |
| Species reactivity: | Human | |
| Calibration range: | ||
| Recovery (accuracy): | ||
| Linearity of dilution: | ||
| Limit of quantitation: | 11.65 ng/mL | |
| Limit of detection: | 7.73 ng/mL | |
| Precision: | ||
| Intra-assay %CV: | 4.0% | |
| Inter-assay %CV: | 1.6% | |
| Stability: | 24 months from manufacture date | |
| Storage: | IMPORTANT: Store Calibrators at -20°C or colder upon receipt. Store remaining reagents at 2-8°C. | |
| Downloads: | IFU MSDS 2012_ENDO_PAPP-A Poster | |
| Description: |
PAPP-A (Pregnancy Associated Plasma Protein A) |
|
| Catalog number: | PAPP-A ELISA, AL-106 | |
| Kit / type: | 96-well (12×8 removable strips) microtiter plate ELISA | |
| Detection: | HRP-based ELISA, colorimetric detection by dual wavelength absorbance at 450 nm and 630 nm as reference filter | |
| Intended use: | Unless otherwise stated in our catalog or other company documentation accompanying Ansh Labs product(s), the products are intended for research use only, and not for in vitro diagnostic purposes, or therapeutic uses of any type involving consumption or application to humans or animals. | |
| Assay note(s): | Quantitative two-step sandwich type immunoassay. Apprpriate sample dilution should be performed same days as assay. Total 3 hour incubation time at room temperature. | |
| Sample type(s): | Serum, diluted 1:100 | |
| Sample size: | 50 μL of diluted specimens, 10 μL specimen into 990 μL assay buffer | |
| Analytical Specificity: | The antibody pair used in the PAPP-A kits measures dimeric PAPP-A and PAPP-A/proMBP complex approximately in equimolar concentration and does not cross-react with PAPP-A2 and MMP-9 when tested at 50 ng/mL. | |
| Species reactivity: | Human | |
| Calibration range: | ||
| Recovery (accuracy): | ||
| Linearity of dilution: | ||
| Limit of quantitation: | 36 ng/mL | |
| Limit of detection: | 10.10 ng/mL | |
| Precision: | ||
| Intra-assay %CV: | 2.2% | |
| Inter-assay %CV: | 2.1% | |
| Stability: | 18 months from manufacture date | |
| Storage: | IMPORTANT: Store Calibrators at -20°C or colder upon receipt. Store remaining reagents at 2-8°C. | |
| Downloads: | IFU MSDS | |
| Description: |
PAPP-A (Pregnancy Associated Plasma Protein A) |
|
| CE | ||
| Catalog number: | AnshLite™ PAPP-A CLIA, AL-206-i | |
| Kit / types: | 96-well (12×8 removable strips) microtiter plate CLIA | |
| Detection: | HRP-based enhanced chemiluminescence, utilizing our proprietary AnshLite™ luminogenic substrate | |
| Intended use: | FOR SALE OUTSIDE THE UNITED STATES. The PAPP-A kit provides materials for the quantitative measurement of PAPP-A in serum and other biological fluids. This assay is intended for in vitro diagnostic use. | |
| Assay note(s): | Quantitative two-step sandwich type immunoassay. Apprpriate sample dilution should be performed same days as assay. Total 3 hour incubation time at room temperature. | |
| Sample type(s): | Serum, diluted 1:100 | |
| Sample size: | 50 μL of diluted specimens, 10 μL specimen into 990 μL assay buffer, | |
| Analytical Specificity: | The antibody pair used in the PAPP-A kits measures dimeric PAPP-A and PAPP-A/proMBP complex approximately in equimolar concentration and does not cross-react with PAPP-A2 and MMP-9 when tested at 50 ng/mL. | |
| Species reactivity: | Human | |
| Calibration range: | ||
| Recovery (accuracy): | ||
| Linearity of dilution: | ||
| Limit of quantitation: | 11.65 ng/mL | |
| Limit of detection: | 7.73 ng/mL | |
| Precision: | ||
| Intra-assay %CV: | 4.0% | |
| Inter-assay %CV: | 1.6% | |
| Stability: | 24 months from manufacture date | |
| Storage: | IMPORTANT: Store Calibrators at -20°C or colder upon receipt. Store remaining reagents at 2-8°C. | |
| Downloads: | IFU MSDS 2012_ENDO_PAPP-A Poster | |
| Description: |
PAPP-A (Pregnancy Associated Plasma Protein A) |
|
| CE | ||
| Catalog number: | PAPP-A ELISA, AL-106-i | |
| Kit / types: | 96-well (12×8 removable strips) microtiter plate ELISA | |
| Detection: | HRP-based ELISA, colorimetric detection by dual wavelength absorbance at 450 nm and 630 nm as reference filter | |
| Intended use: | FOR SALE OUTSIDE THE UNITED STATES. The PAPP-A kit provides materials for the quantitative measurement of PAPP-A in serum and other biological fluids. This assay is intended for in vitro diagnostic use. | |
| Assay note(s): | Quantitative two-step sandwich type immunoassay. Apprpriate sample dilution should be performed same days as assay. Total 3 hour incubation time at room temperature. | |
| Sample type(s): | Serum, diluted 1:100 | |
| Sample size: | 50 μL of diluted specimens, 10 μL specimen into 990 μL assay buffer, | |
| Analytical Specificity: | The antibody pair used in the PAPP-A kits measures dimeric PAPP-A and PAPP-A/proMBP complex approximately in equimolar concentration and does not cross-react with PAPP-A2 and MMP-9 when tested at 50 ng/mL. | |
| Species reactivity: | Human | |
| Calibration range: | ||
| Linearity of dilution: | ||
| Limit of quantitation: | 36 ng/mL | |
| Limit of detection: | 10.10 ng/mL | |
| Precision: | ||
| Intra-assay %CV: | 2.2% | |
| Inter-assay %CV: | 2.1% | |
| Stability: | 18 months from manufacture date | |
| Storage: | IMPORTANT: Store Calibrators at -20°C or colder upon receipt. Store remaining reagents at 2-8°C. | |
| Downloads: | IFU MSDS | |